58 Exam-based calculations
4 Using the touchpad, trace the waveform.
To make a correction, select Undo on-screen,
backtrack with the touchpad, or press the
BACKSPACE key.
5 Press the
SET key to complete the trace.
6 Save the calculation. (See “To save a
calculation” on page 47.)
For information on the automatic trace tool, see
“To trace automatically (Doppler)” on
page 46.
To calculate Right Ventricular Systolic
Pressure (RVSP)
1 On a frozen Doppler spectral trace, press the
CALCS key.
2 From the calculations menu, select TV and then
select TRmax.
3 Position the caliper. (See “Working with
calipers” on page 43.)
4 Save the calculation. (See “To save a
calculation” on page 47.)
5 To adjust the RA pressure, see “To delete a
vascular or cardiac measurement” on
page 72.
Changing the RA pressure from the default 5
affects the RVSP calculation in the patient
report.
To calculate Pressure Half Time (PHT) in
MV, AI, or TV
1 On a frozen Doppler spectral trace, press the
CALCS key.
2 From the calculations menu, select MV, AV, or
TV, and then select PHT.
3 Position the first caliper at the peak, and press
the
SELECT key.
A second caliper appears.
4 Position the second caliper:
In MV, position the caliper along the EF
slope.
In AV, position the caliper at the end
diastole.
5 Save the calculation. (See “To save a
calculation” on page 47.)
To calculate Proximal Isovelocity Surface
Area (PISA)
The PISA calculation requires a measurement in 2D,
a measurement in Color, and two measurements in
Doppler spectral trace. After all measurements are
saved, the result appears in the patient report.
1 Measure from Ann D (2D):
a On a frozen 2D image, press the
CALCS key.
b From the calculations menu, locate PISA,
and then select Ann D.
c Position the calipers. (See “Working with
calipers” on page 43.)
d Save the calculation. (See “To save a
calculation” on page 47.)
2 Measure from Radius (Color):
a On a frozen Color image, press the
CALCS
key.
b From the calculations menu, select Radius.
c Position the calipers.
d Save the calculation.
3 On a frozen Doppler spectral trace, press the
CALCS key.
4 Do the following to measure from MR VTI and
again to measure from MV VTI (Doppler):
a From the calculations menu, select PISA and
then select MR VTI or MV VTI.
b Position the caliper at the start of the
waveform, and press the
SELECT key to start
the trace.
c Using the touchpad, trace the waveform.