92 Cleaning and disinfecting
To clean the LCD screen
Dampen a clean, non-abrasive, cotton cloth with
an ethanolic-based cleaner, and wipe the screen
clean.
Apply the cleaner to the cloth rather than the
surface of the screen.
To clean and disinfect system surfaces
1 Turn off the system.
2 Disconnect the system from the power supply,
or remove it from the mini-dock or docking
system.
3 Clean the exterior surfaces using a soft cloth
lightly dampened in a mild soap or detergent
cleaning solution to remove any particulate
matter or body fluids.
Apply the solution to the cloth rather than the
surface.
4 Mix the disinfectant solution compatible with the
system, following disinfectant label instructions
for solution strengths and disinfectant contact
duration.
5 Wipe surfaces with the disinfectant solution.
6 Air dry or towel dry with a clean cloth.
Cleaning and disinfecting transducers
To disinfect the transducer and its cable, use the
immersion method or the wipe method.
Caution:
When you clean the system,
ensure that the solution does not
get inside the system controls or
the battery compartment.
Caution:
Do not scratch the LCD screen.
WARNING:
To avoid electrical shock, before
cleaning, disconnect the
transducer from the system.
WARNING:
To avoid injury, always use
protective eyewear and gloves
when performing cleaning and
disinfecting procedures.
WARNING:
To avoid infection, ensure that the
solution expiration date has not
passed.
Caution:
Transducers must be cleaned after
every use. Cleaning transducers is
necessary prior to effective
disinfection. Ensure that you follow
the manufacturer's instructions
when using disinfectants.
Caution:
Do not use a surgeon's brush when
cleaning transducers. Even the use
of soft brushes can damage a
transducer. Use a soft cloth.
Caution:
Using a non-recommended
cleaning or disinfection solution,
incorrect solution strength, or
immersing a transducer deeper or
for a longer period of time than
recommended can damage or
discolor the transducer and void
the transducer warranty.
Caution:
Do not allow cleaning solution or
disinfectant into the transducer
connector.