Chapter 7: Safety 101
Safety
Clinical safety
WARNING:
Do not use a damaged battery.
WARNING:
Do not solder a battery.
WARNING:
The polarity of the battery terminals are fixed and cannot be switched or reversed.
Do not force the battery into the system.
WARNING:
Do not connect the battery to an electrical power outlet.
WARNING:
Do not continue recharging the battery if it does not recharge after two successive
six hour charging cycles.
WARNING:
Do not ship a damaged battery without instructions from SonoSite Technical
Support. (See “SonoSite Technical Support” on page ix.)
WARNING:
If the battery leaks or emits an odor, remove it from all possible flammable sources.
WARNING:
Periodically, check to make sure that the battery charges fully. If the battery fails to
charge fully, replace it.
Caution:
To avoid the battery becoming damaged and causing equipment damage, observe
the following precautions:
Do not immerse the battery in water or allow it to get wet.
Do not put the battery into a microwave oven or pressurized container.
If the battery emits an odor or heat, is deformed or discolored, or in any way
appears abnormal during use, recharging or storage, immediately remove it and
stop using it. If you have any questions about the battery, consult SonoSite or your
local representative.
Store the battery between -20°C (-4°F) and 60°C (140°F).
Use only SonoSite batteries.
Do not use or charge the battery with non-SonoSite equipment. Only charge the
battery with the system.
WARNING:
Non-medical (commercial) grade peripheral monitors have not been verified or
validated by SonoSite as being suitable for diagnosis.
WARNING:
To avoid the risk of a burn hazard, do not use the transducer with high frequency
surgical equipment. Such a hazard may occur in the event of a defect in the high
frequency surgical neutral electrode connection.