Chapter 7: Safety 103
Safety
Electromagnetic compatibility
The ultrasound system has been tested and found to comply with the electromagnetic compatibility
(EMC) limits for medical devices to IEC 60601-1-2:2001. These limits are designed to provide
reasonable protection against harmful interference in a typical medical installation.
WARNING:
The Edge ultrasound system should not be used adjacent to or stacked with other
equipment. If such use occurs, verify that the Edge ultrasound system operates
normally in that configuration.
Caution:
Medical electrical equipment requires special precautions regarding EMC and must
be installed and operated according to these instructions. Portable and mobile RF
communications equipment can affect the ultrasound system. Electromagnetic
interference (EMI) from other equipment or interference sources could result in
performance disruption of the ultrasound system. Evidence of disruption may
include image degradation or distortion, erratic readings, equipment ceasing to
operate, or other incorrect functioning. If this occurs, survey the site to determine
the source of disruption, and take the following actions to eliminate the source(s).
Turn equipment in the vicinity off and on to isolate disruptive equipment.
Relocate or re-orient interfering equipment.
Increase distance between interfering equipment and your ultrasound system.
Manage use of frequencies close to ultrasound system frequencies.
Remove devices that are highly susceptible to EMI.
Lower power from internal sources within facility control (such as paging
systems).
Label devices susceptible to EMI.
Educate clinical staff to recognize potential EMI-related problems.
Eliminate or reduce EMI with technical solutions (such as shielding).
Restrict use of personal communicators (cell phones, computers) in areas with
devices susceptible to EMI.
Share relevant EMI information with others, particularly when evaluating new
equipment purchases which may generate EMI.
Purchase medical devices that comply with IEC 60601-1-2 EMC Standards.
Caution:
To avoid the risk of increased electromagnetic emissions or decreased immunity,
use only accessories and peripherals recommended by SonoSite. Connection of
accessories and peripherals not recommended by SonoSite to the ultrasound
system may result in malfunction of the ultrasound system or other medical
electrical devices in the area. Contact SonoSite or your local representative for a
list of accessories and peripherals available from or recommended by SonoSite.
See the SonoSite accessories user guide.