Chapter 8: Acoustic Output 149
Acoustic Output
(a) This index is not required for this operating mode; value is <1.
(b) This transducer is not intended for transcranial or neonatal cephalic uses.
#No data are reported for this operating condition since the global maximum index value is not reported for
the reason listed. (Reference Global Maximum Index Value line.)
— Data are not applicable for this transducer/mode.
Table 26: Transducer Model:
L38xi/10-5
Operating Mode:
CPD/Color
Index Label M.I.
TIS TIB
TIC
Scan
Non-scan
Non-scan
A
aprt
1A
aprt
>1
Global Maximum Index Value 1.5 1.1 (b)
Associated Acoustic
Parameter
p
r.3
(MPa)
3.30
W
0
(mW)
47.5 —#
min of [W
.3
(z
1
),I
TA.3
(z
1
)] (mW)
z
1
(cm)
z
bp
(cm)
z
sp
(cm)
0.8
d
eq
(z
sp
)(cm)
f
c
(MHz)
4.82 4.82 #
Dim of A
aprt
X(cm) 0.66 #
Y(cm)
0.40 #
Other Information
PD (μsec) 0.544
PRF (Hz) 2885
p
r
@PII
max
(MPa)
3.79
d
eq
@Pll
max
(cm)
Focal Length FL
x
(cm)
1.86 #
FL
y
(cm)
1.50 #
I
PA.3
@MI
max
(W/cm
2
)
605
Operating
Control
Conditions
Control 1: Mode CVD/CPD CVD
Control 2: Exam Type Any Bre
Control 3: 2D Optimization/Depth Any/2.0-
2.5 cm
Any/3.8
cm
Control 4: Color Optimization/PRF
Any/Any
Low/
1323
Control 5: Color Box Position/Size
Any/Any
Any/
Default