WARNING:
To avoid the risk of electrical shock:
• This equipment must be connected only to a supply mains with protective earth.
• Use only properly grounded equipment. Shock hazards exist if the power supply
is not properly grounded. Grounding reliability can be achieved only when
equipment is connected to a receptacle marked “Hospital Only” or “Hospital
Grade” or the equivalent. The grounding wire must not be removed or defeated.
• When using the system in an environment where the integrity of the protective
earth conductor arrangement is in doubt, operate the system on battery power
only without using the power supply.
• Do not let the bar code scanner or external mouse touch the patient.
• Do not touch any of the following:
• The power supply and the patient at the same time
• The ungrounded signal input/output connectors on the back of the ultrasound
system
• The system battery contacts (inside the battery compartment)
• The system transducer connector when the transducer or Triple Transducer
Connect (TTC) is disconnected
• The system transducer connector on the TTC if no transducers are connected
• Do not connect the system’s power supply or a docking system to a multiple
portable socket outlet (MPSO) or extension cord.
• Before using the transducer, inspect the transducer face, housing, and cable. Do
not use the transducer if the transducer or cable is damaged.
• Always disconnect the power supply from the system before cleaning the system.
• Do not use any transducer that has been immersed beyond the specified cleaning
or disinfection level. See Chapter 6, “Troubleshooting and Maintenance.”
• Use only accessories and peripherals recommended by SonoSite, including the
power supply. Connection of accessories and peripherals not recommended by
SonoSite could result in electrical shock. Contact SonoSite or your local
representative for a list of accessories and peripherals available from or
recommended by SonoSite.
• Use commercial grade peripherals recommended by SonoSite on battery power
only. Do not connect these products to AC mains power when using the system
to scan or diagnose a patient/subject. Contact SonoSite or your local
representative for a list of the commercial grade peripherals available from or
recommended by SonoSite.
WARNING:
To avoid the risk of electrical shock and fire hazard:
• Inspect the power supply, AC power cords, cables, and plugs on a regular basis.
Ensure that they are not damaged.
• The power cord set that connects the power supply of the ultrasound system or
the stand to mains power must only be used with the power supply or docking
system, and cannot be used to connect other devices to mains power.