Chapter 4: Measurements and Calculations 51
Measurements
Volume flow calculations
The following table shows the measurements
required to complete the volume flow calculation.
For definitions of acronyms, see “Glossary” on
page 167.
Both a 2D and a Doppler measurement are required
for the volume flow calculation. For the 2D
measurement, you can do either of the following:
Measure the diameter of the vessel. This
approach is more precise. The measurement
overrides the gate size.
Use the gate size. If you do not measure the
diameter of the vessel, the system automatically
uses the gate size and “(gate)” appears in the
calculation results.
The Doppler sample volume should completely
insonate the vessel. You can measure either the
time average mean or time average peak. To
specify the live trace setting, see “Presets
setup” on page 22.
Consider the following factors when performing
volume flow measurements:
Users should follow current medical practice for
volume flow calculation applications.
The accuracy of the volume flow calculation
largely depends on the user.
The factors identified in the literature that affect
the accuracy are as follows:
Using the diameter method for 2D area
Difficulty ensuring uniform insonation of the
vessel.
WARNING:
To avoid incorrect calculations,
verify that the patient information,
date, and time settings are
accurate.
WARNING:
To avoid misdiagnosis or harming
the patient outcome, start a new
patient information form before
starting a new patient exam and
performing calculations. Starting a
new patient information form
clears the previous patient’s data.
The previous patient’s data will be
combined with the current patient
if the form is not first cleared. See
“To create a new patient
information form” on page 36.
Transducer Exam Types
C11x Abdomen, Vascular
C60x Abdomen
HFL38x Vascular
L25x Vascular
L38x Vascular
L38xi Vascular
P10x Abdomen
P21x Abdomen
SLAx Vascular
Volume Flow Calculations
Menu
Heading
Measurement
(Imaging Mode)
Calculation
Result
Vol Flow D (2D)*
TAM or TAP
(Doppler)
VF (Volume
Flow ml/min)
* Required if measuring the diameter instead of using the
gate size