Chapter 8: Performance Testing 59
Chapter 8: Performance Testing
Overview
To obtain 2D images, SonoSite recommends using the Gammex 403GS Soft Tissue Phantom or the Gammex 413A
Multipurpose Phantom. A .7db/cm phantom is recommend but not required.
Some features and capabilities are optional and therefore may be unavailable to test.
Recommend Test Equipment
SonoSite ultrasound system under test
C60x/5-2 MHz transducer
P21x/5-1 MHz transducer
Gammex 403 GS Multipurpose Phantom, 413A Soft Tissue Phantom, or equivalent.
•Acoustic gel
Setting Up Performance Tests
WARNING:
Critical Test Function — A failure of the system functions tested in this section could affect
safety or effectiveness of the system adversely. While performing the steps in this section, verify
that the images on the system display and on the external monitor are acceptable.
Set up
Performance
Tests
1 Attach the C60x/5-2 MHz transducer to the system.
2 Select Gen for optimization and OB for exam type.
3 Couple the transducer to the phantom, adjusting gain settings and transducer for a proper
phantom image (e.g., pins are high-level echoes positioned in straight lines; cysts are
sonolucent, edges are sharp, and graphite particles of the phantom are mid-grays).