48 Chapter 6: Replacement Procedures
Power Supply PCBA Replacement
Required Part
P08850 Service Assembly, Power Supply, M-Turbo (Compatible with the EDGE system)
Power Supply Removal
1 Remove the 7 long screws and 2 short screws holding the Power Supply Shield. Please note the location of
the 2 short screws. See Figure 6.11.
Figure 6.11 Power Supply Shield
2 Gently lift the Power Supply PCBA away from the Main PCBA as shown in Figure 6.12. (Connector in upper left
corner/removing the Power Supply PCB without the shield is difficult)
3 Install the new Power Supply PCBA by reversing steps 1-2.
Figure 6.12 Power Supply PCBA
Power Supply
shield
Long Screws (7x)
Short Screws (2x)
Power Supply connector
underneath mates Power
Supply PCB to the Main
PCBA.
Power Supply PCBA